In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.
CVSS Details
- CVSS 3.1 Base Score: 4.3
- CVSS 3.1 Vector: (CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L)