Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted, stolen, or modified, or the Paceart Optima system being used for further network penetration via network connectivity.
With curated Threat Intelligence, you can see which vulnerabilities truly put you at risk, prioritize what matters most, and act before attackers do.
Explore Intelligence Hub