An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.
CVSS Details
- CVSS 3.1 Base Score: 8.1
- CVSS 3.0 Vector: (CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H)